X-ray HV · MRI Gradient · Defibrillator · Ventilator · IEC 60601

Resistors for Medical & Imaging

Medical imaging and life-support equipment subject resistors to combined HV stress and ME (medical electrical) safety requirements. X-ray generator anode supplies use 70–150 kV resistor dividers with TCR < 25 ppm/°C and voltage coefficient < 1 ppm/V. MRI gradient amplifiers dissipate up to 30 kW peak in damping resistors during slew-rate transitions. External defibrillators store 360 J in a 32 µF capacitor and use a current-limiting / discharge resistor of 50 Ω rated at 10 kV impulse. Ventilator blower drives, ECMO pumps and surgical lasers also rely on snubber and brake resistors. Compliance includes IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-44 (CT), IEC 60601-2-33 (MRI) and ISO 13485 quality system. Hongyi supplies ISO 13485-traceable parts with full risk-management file per ISO 14971 to FDA QSR and EU MDR submissions.

Key Requirements

Key Requirements

  • IEC 60601-1 ed.3.2 working voltage and MOPP (means of patient protection) compliance
  • X-ray HV divider TCR < 25 ppm/°C and VCR < 1 ppm/V for stable kVp accuracy
  • Defibrillator pulse rating 10 kV / 50 µs / 360 J single-shot capability
  • ISO 13485 design-history file with biocompatibility cross-reference if surface-contact
  • EMC compliance to CISPR 11 Group 1 Class A radiated and conducted
How to Select

How to Select

  1. Specify HV divider with paired matched-pair TCR tracking ≤ 5 ppm/°C for kVp ratiometric stability
  2. Confirm partial-discharge inception ≥ 1.2× rated voltage per IEC 60270
  3. Select non-magnetic materials (no nickel cores) for MRI bore-side parts
  4. Validate single-pulse failure mode is open-circuit (not flash-over) for defibrillator safety
  5. Demand UDI labeling and lot traceability for FDA / MDR post-market surveillance

Frequently asked questions

Are your medical-grade resistors ISO 13485 traceable?
Yes. Our medical resistor line is built in a segregated cell operating under ISO 13485:2016 QMS. Each lot carries a Device Master Record reference, raw-material certificates of analysis, in-process IQC/IPQC records, and a Certificate of Conformity tied to the customer's part number. The DHF can be audited remotely or on-site as part of MDSAP or FDA inspections.
Can you supply non-magnetic resistors for MRI bore-side gradient amplifiers?
Yes. Our non-magnetic series replaces nickel-plated terminations and ferrous bracketry with copper / brass / aluminum and 316L stainless. Residual permeability is < 1.005 measured per ASTM A342 at 3 T. Typical use is in MRI room cabinets carrying 30 A gradient drive at 800 V slew. Specify part suffix /NM when ordering.
How do I verify the pulse rating for a defibrillator discharge resistor?
Compute pulse energy E = ½ × C × V². A 360 J monophasic defibrillator at 5 kV with 32 µF capacitor gives 400 J theoretical, derated 80%. Our 50 Ω 10 kV pulse resistor is tested at 500 J single-shot, 100 shots/min for 60 minutes, with R-shift < 0.5%. Datasheet includes 10/1000 µs impulse curve and IEC 60601-2-4 Figure 201.102 compliance evidence.
Do you provide an ISO 14971 risk-management contribution for our submission?
Yes. We provide a component-level hazard analysis covering failure modes (open, short, drift, surface breakdown), severity / probability scoring, mitigation evidence (X-ray inspection, 100% high-pot, conformal coating), and a residual-risk statement. This document slots directly into your device hazard analysis with cross-references to design-output specifications.
Sample Request

Sample Request

Standard samples are provided free of charge in reasonable quantities (shipped within 2–3 business days). Custom samples are delivered within 2–4 weeks. Please use the form on the right to tell us:

  • Company and contact details
  • Target part number or key specs (power / resistance / package)
  • Application scenario and estimated annual volume
  • Shipping address and expected delivery date
📧 You can also email us directly at [email protected] / 13650089870